14.
Buffer sets
SevenEasy pH Meter automatically corrects for setting temperature using the values shown in the
tables:
Buffer Set 1 (ref. 25 °C) MT US
5
7.09
10
7.06
15
7.04
20
7.02
25
7.00
30
6.99
35
6.98
40
6.97
45
6.97
50
6.97
Buffer Set 2 (ref. 25 °C) MT Europa
5
7.09
10
7.06
15
7.04
20
7.02
25
7.00
30
6.99
35
6.98
40
6.97
45
6.97
50
6.97
Buffer Set 3 (ref. 20 °C) Merck
5
7.07
10
7.05
15
7.02
20
7.00
25
6.98
30
6.98
35
6.96
40
6.95
45
6.95
50
6.95
Buffer Set 4 (ref. 25 °C) Japan JIS
5
6.95
10
6.92
15
6.90
20
6.88
25
6.86
30
6.85
35
6.84
40
6.84
45
6.83
50
6.83
METTLER TOLEDO SevenEasy
4.00
10.25
1.67
4.00
10.18
1.67
4.00
10.12
1.67
4.00
10.06
1.68
4.00
10.01
1.68
4.01
9.97
1.68
4.02
9.93
1.69
4.03
9.89
1.69
4.04
9.86
1.70
4.06
9.83
1.71
4.01
9.45
2.02
4.00
9.38
2.01
4.00
9.32
2.00
4.00
9.26
2.00
4.01
9.21
2.00
4.01
9.16
1.99
4.02
9.11
1.99
4.03
9.06
1.98
4.04
9.03
1.98
4.06
8.99
1.98
4.04
9.16
2.01
12.41
4.02
9.11
2.01
12.26
4.01
9.05
2.00
12.10
4.00
9.00
2.00
12.00
4.01
8.95
2.00
11.88
4.01
8.91
2.00
11.72
4.01
8.88
2.00
11.67
4.01
8.85
2.00
11.54
4.01
8.82
2.00
11.44
4.00
8.79
2.00
11.33
4.00
9.40
1.67
4.00
9.33
1.67
4.00
9.28
1.67
4.00
9.22
1.68
4.01
9.18
1.68
4.02
9.14
1.68
4.02
9.10
1.69
4.04
9.07
1.69
4.05
9.04
1.70
4.06
9.01
1.71
11.72
11.54
11.36
11.18
11.00
10.82
10.64
10.46
10.28
10.10
Buffer sets
17