Technical Data:

Measurement system: Oscillometric
Measurement range: 40 – 240 mmHg, 40 – 180 Puls/min.
Cuff pressure: 0 – 320 mmHg
Memory store: 1 measurement
Display: LCD
Working conditions:
Environmental temperature: Room temperature 10 – 40°C
Relative humidity 10 – 85 %
Storing conditions:
Environmental temperature - 5°C to + 50°C
Relative humidity 85 % maximum
Power supply:
DC 6 V (batteries 4 x 1,5 Mignon IEC LR 6, alkaline manganese)
Alternative special option: Main unit DC 6 V, Order-No. 410-
7-153 (Polarity: outside MINUS, inside PLUS)
Battery check: Symbol display in LCD window
Weight: 600 g without batteries
Dimensions (B x H x T):
150 mm x 160 mm x 120 mm
Classification: Class II (
Clinical Test (DIN 58130):
Accuracy complies with EN 1060 part 3
Maximum deviation of cuff pressure: ± 3 mmHg
Maximum deviation of pulse rate: ± 5 %