4 Operation

4.1 Electrical Connections

SLP85xD load cells have the following interfaces:
Power supply for the load cell
The load cells work with 12 to 24 V DC nominal range (10 to 30 V DC). Input current: 0.84 A
Use a stable power supply with no voltage fluctuations.
If voltage fluctuations cannot be prevented, use a voltage regulator to deliver a constant voltage value to
the load cell.
The power supply must be approved by the respective national test center of the country in which the
load cell will be used.
The power is supplied to the load cell via pins A and B in the communication connector.
Communication
Connector (2)
B
C
D
Digital inputs / outputs
Digital inputs
Number
Permissible input voltage 0 to +30 V
Low level
High level
Input resistance

4.2 Environmental Conditions

SLP85xD load cells can be operated within the following environmental conditions:
Temperature Range

4.3 Ingress Protection

SLP85xD load cells are hermetically sealed and have an IP rating of IP68 and IP69k
SLP85xD
2
1
Pin
Signal
Industrial Ethernet
A
UB1
B
GND1
A
G
H
F
E
3
<1.5 V
>10 V
>1.2 kOhms
Compensated
Operating
Safe storage
1
IO connector, SLP854D
only
2
Communication
connector
RS232 service interface
Profibus DP
UB1
GND1
Digital outputs
Number
External supply voltage
Max. current of one
output
Accumulated current of
all outputs
Voltage drop
–10 ... +40 °C
–10 ... +50 °C
–20 ... +80 °C
3 digital inputs
5 digital outputs
Bus interface
(Profibus DP or
Ethernet IP or Profinet IO)
Service interfae (RS232)
Cable color
Red
Black
5
+11 to +30 V
<0.5 A
<1.2 A
<1 V
Operation