4 Operation

4.1 Electrical connections

WMC weigh modules have a standard RS232C interface for the communication with the control systems.
Additionally, a second optional interface can be added, which can be either a second RS232C interface or
an Ethernet TCP/IP interface.
Interface descriptions
Interface type
Interface update rate (max.)
1)
38400 baud only possible:
• Weigh module without terminal, or
• Weigh module with terminal, only via the optional RS232C interface.
Power supply for the weigh module
Input voltage
Input current
Use a stable power supply with no voltage fluctuations.
If voltage fluctuations cannot be prevented, use a voltage regulator to deliver a constant voltage value to
the load cell.
The power supply must be approved by the respective national test center of the country in which the
weigh module will be used.
Always operate the weigh module with the standard AC adapter delivered in the scope of delivery. The AC
adapter is suitable for use only with the following input characteristics:
100 – 240 V AC, -15%/+10%, 50/60 Hz, 0.5A
Note
Check whether your local power supply falls within this range. If it does not, DO NOT connect the electronic
unit or the AC adapter to your power supply and contact your local METTLER TOLEDO office.
If you use an alternative AC adapter other than the one in the standard scope of delivery, make sure that it is
a tested AC adapter with SELV output current and ensure correct polarity.

4.2 Environmental conditions

WMC weigh modules can be operated within the following environmental conditions:
Temperature Range
Relative air humidity
Height above mean sea level
Warm-up time

4.3 Ingress protection

IP rating of the WMC weigh modules:
8
Operation​ ​
RS232C, bidirectional, full duplex, 600 to 38'400
(baud rate can be selected using interface commands)
23 weight values per second (with terminal)
92 weight values per second (without terminal)
12 V DC (nominal)
2 ADC
Operating /
Compensated
Allowable ambient
+10 ... +30 °C
+5 ... +40 °C
Max. 80 % at 31 °C, decreasing linearly to 50 %
at 40 °C, non-condensing
Max. 4'000 m (13'330 ft)
At least 60 minutes after power up
At least 60 minutes after the weigh module has
been connected to the power supply; the weigh
module can be used immediately if it is turned on
from standby mode (when operated with a terminal
connected).
1)
bps
Weigh Modules