Table 2
For all ME EQuIPMENT and ME sYsTEMs
Guidance and manufacturer's declaration – electromagnetic immunity
The FT 100 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FT 100 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
6 kV contact
(ESD)
8 kV air
IEC 61000-4-2
Electrical fast transient/
2 kV for power supply lines
burst
1 kV for input/output lines
IEC 61000-4-4
Surge
1 kV line(s) to line(s)
IEC 61000-4-5
2 kV line(s) to earth
interruptions and voltage
< 5 % UT (> 95 % dip in UT)
variations on power supply
for 0.5 cycle
input lines
40 % UT (60 % dip in UT) for
IEC 61000-4-11
5 cycles
70 % UT (30 % dip in UT) for
25 cycles
< 5 % UT (> 95 % dip in UT)
for 5 sec
Power frequency (50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Compliance level
Electromagnetic environment – guidance
6 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with
8 kV air
synthetic material, the relative humidity should be at least 30 %.
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment.
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment.
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the FT 100 requires continued operation during
power mains interruptions, it is recommended that the FT 100 be powered
from an uninterruptible power supply or a battery.
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
173