8. Forehead measurement

Beurer tip
Remember:
• Physical activity, increased perspiration on the forehead, taking vasoconstrictive medication and skin
irritations can distort the measurement.
• The forehead, or the temples, must be free from perspiration and cosmetics.
• We recommend that you take three measurements on the forehead. If the three measurements are
different, use the highest value.
• If the forehead area is covered in hair, sweat or dirt, clean the area and then wait for 10 minutes to
improve the precision of the measurement.
• It is important that you keep the thermometer still during measurement. Moving the thermometer during
measurement may cause imprecise measurements.
• To switch on the thermometer, briefly press the
short self-test, the device is ready for forehead temperature measurement.
The device is in forehead temperature mode. You can tell this from the
symbol on the display.
• Hold the thermometer 4 to 6 cm in front of the forehead measuring point
(Fig. 10). Press the
uring distance of 4-6 cm.
/ sCAN button. The display shows the required meas-
/ sCAN button. After a