Specifications
3.1 Measurement Specifications
Common measurement specifications
Guaranteed
1 year (Opening and Closing of the Sensor: Maximum 10000
accuracy period
times)
Accuracy guaran-
23°C±5°C (73°F±9°F) 80%RH or less (non-condensation)
tee for tempera-
ture and humidity
Temperature
-10 to 50°C Measurement accuracy x 0.1/°C (except 23°C±5°C)
characteristics
Maximum rated
6 00 VA C me a su re me nt ca t eg or y I V ( an t ic ip at ed t r an -
voltage to earth
s ien t o v er vo lta ge 80 00 V)
Resistance measurement specifications
Guaranteed
No reactance component, no noise current
accuracy
conditions
Measurement
Analog synchronous detection method (effective resistance mea-
Method
surement)
Injected signal
Approx. 2.4 kHz
frequency
Injected voltage
Approx. 9.0 mV (with load open)
level
0.02 Ω to 1600 Ω
Effective
measuring range
Zero suppression Less than 0.02 Ω
Greater than 1600 Ω
Overrange
Measurement
Filter: OFF/ON
response time
3.1 Measurement Specifications
Approx. 3 sec /Approx. 9 sec.
Chapter 3
55