7 Reprocessing
42
7
Reprocessing
7.1
Automated cleaning
The cleaning equipment used must be approved by its
manufacturer for the cleaning of dental instruments.
For further details, refer to the operating instructions supplied with
the unit.
NOTE! Do not carry out automated cleaning of the quick
couplings.
ü
The high-speed handpiece is conditioned with a cleaning
device.
1.
Check whether the high-speed handpiece is clean after
conditioning under good lighting (min. 500 lux) and color
rendering index (min. 80 Ra).
2.
If they are still dirty, repeat the process.
Operating instructions Midwest Stylus Plus / Midwest Tradition
67 27 148 D3673.201.09.01.02