Ä
For further processing, the high-speed handpiece should
be dry and free of residue.
3.
Blow the high-speed handpiece out with max. 3 bar.
4.
Maintain mechanical parts manually  [→ 47].
5.
Maintain the push button chuck manually  [→ 50].
6.
Pack the high-speed handpiece in packaging material suitable
for sterilization and storage. e.g. paper/plastic composite
packaging.
7.
Conduct sterilization  [→ 54].
7 Reprocessing