Operating instructions Midwest Stylus Plus / Midwest Tradition
67 27 148  D3673.201.09.01.02
Procedure
ü
The spray adapter is disinfected.
ü
The O-rings on the spray adapter are intact.
1.
Fit the spray adapter onto the nozzle of the spray can.
2.
Insert the high-speed handpiece as far as it will go and hold it
in place.
3.
Spray into the high-speed handpiece for 1-2 seconds.
IMPORTANT: Hold the spray can upright.
4.
Wipe away any spray that runs out with a clean, lint-free cloth.
5.
Repeat the process until the spray that runs out is clear.
7 Reprocessing