7 Reprocessing
46
Operating instructions Midwest Stylus Plus / Midwest Tradition
7.
Sterilize the high-speed handpiece  [→ 54].
8.
Disinfect the quick coupling with disinfectant wipes.
Observe the manufacturer's instructions for using the
instrument disinfectant.
9.
Wipe the disinfectant off with a clean, lint-free cloth.
10. Purge the individual connections of the quick coupling with
compressed air for at least 10 seconds.
67 27 148  D3673.201.09.01.02