SoftCloud® Ace, SoftCloud® Pro

6.2.3 Disinfection Instructions

IMPORTANT!
– Only use disinfectants and methods approved
by your local infection control institution and
follow your local infection control policy.
– Additional information on recommended
disinfectants can be found on
https://vah-online.de/en/for-users.
Disinfecting the Control Unit
1. Wipe down all generally accessible surfaces with a cloth
and disinfectant.
2. Allow the product to air-dry.
Disinfecting the Air Cells
IMPORTANT!
– Do not disconnect the pipes from individual air
cells.
– Do not machine wash or dry the air cells.
1. Disconnect air cells from the base by unfastening the
press studs at each end.
2. Disconnect air pipes from main air hoses.
3. Slide each cell out from the cell straps.
4. Swab with a cloth damped with warm water containing
an appropriate disinfectant (dependent on infection
control protocols and local market requirements).
5. Dry thoroughly with a soft cloth before refastening.

6.3 Service

IMPORTANT!
– The service indicator will be activated when the
control unit has been running for approximately
9.000 hours, indicating that service is needed.
IMPORTANT!
– Do not attempt to repair or service the control
unit yourself! Always contact an authorised
service technician.
14

6.4 Replacing the air filter

IMPORTANT!
Good filter maintenance is critical to maintain
your mattress system in optimal operating
condition. Failure to keep the filters clean will
result in system downtime and increase repair
costs. It is recommended that the air filter is
replaced annually. Replacement air filters are
available from an Invacare Service Centre.
1. Turn off the power supply to the control unit.
2. Disconnect the power cord and air hoses.
3. Place the control unit on a flat surface with its back
panel uppermost (place soft cloth under unit to prevent
scratches).
4. Carefully remove the air filter shroud. Remove and
discard the filter and attach with new filter.
5. Re-attach the air filter shroud to the control unit.
The control unit is now ready for re-connection.

6.5 Replacing fuse

CAUTION!
– Ensure the replacement of fuses is carried out
in accordance with local legislation.
– Always use the same fuse rating when replacing
a defect fuse.
1. Turn off the power supply to the control unit.
2. Remove the power cord from the electrical socket on
the side of the control unit.
3. Insert a flat screwdriver into the small cutout above the
power cord inlet.
4. Pull the screwdriver carefully towards to open the plastic
carrier.
5. Remove and discard the "blown" fuse.
6. Insert a new fuse and close the plastic carrier.
7. Reverse steps four to one in order to reinstate the fuse
carrier.
See also section "Activating the control unit" for
more information.