CAUTION
Dirty thread in screw connection
Injuries due to falling when thread loosens, breaks or tears out
► Use a degreasing cleaner (e.g. Acetone) to clean the thread before
each mounting.
► Observe the alignment and assembly instructions.
Required tools and materials:
>
710D4 Torque Wrench, 636K13 Loctite® 241
1) If a socket adapter is installed at the shuttle lock: remove the socket
adapter.
2) Apply Loctite® to the thread.
3) Position the slider plate at the shuttle lock and screw it in using the
countersunk head screws (see fig. 2, see fig. 3) (mounting torque: 12
Nm).
4) Apply Loctite® to the thread of the countersunk head screws.
5) Position the socket adapter at the slider plate and then use the counter­
sunk head screws to fasten it (see fig. 4, see fig. 5) (mounting torque:
12 Nm).
6 Maintenance
► A visual inspection and functional test of the prosthetic components
should be performed after the first 30 days of use.
► Inspect the entire prosthesis for wear during normal consultations.
► Conduct annual safety inspections.

7 Legal information

All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
7.1 Liability
The manufacturer will only assume liability if the product is used in accord­
ance with the descriptions and instructions provided in this document. The
manufacturer will not assume liability for damage caused by disregarding the
information in this document, particularly due to improper use or unauthor­
ised modification of the product.
7.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745on medical
devices. The CE declaration of conformity can be downloaded from the man­
ufacturer's website.