- in the Tube Adapter 3R41=1:
- in the Tube Adapter 2R48:
Secure the screws by means of 636K13 Loctite.
Attention!
To eliminate noise in the cosmetic foam cover, use 519L5 Silicone
Spray. Do not use talcum powder! Talcum powder reduces the
lubrication of the mechanical parts, which may lead to a malfunc-
tion and thus increase the risk of falling. Using this medical product
after application of talcum powder will render all claims against
Ottobock null and void.
Apply Silicone Spray 519L5 directly on the friction surfaces of the
cosmetic foam cover!
5 Servicing
Due to the extreme stresses that children place on the components, regular
inspections and servicing by the prosthetist are necessary at pre- determi-
ned intervals. We recommend that the components be inspected every
3 months. Function should be tested and the tube adapters examined for
any evidence of deformities or damage.
6 Disposal
In some jurisdictions it is not permissible to dispose of the product with un-
sorted household waste. Improper disposal can be harmful to health and the
environment. Observe the information provided by the responsible authorities
in your country regarding return, collection and disposal procedures.
7 Legal information
All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
7.1 Liability
The manufacturer will only assume liability if the product is used in accor-
dance with the descriptions and instructions provided in this document. The
manufacturer will not assume liability for damage caused by disregard of
this document, particularly due to improper use or unauthorised modification
of the product.
14 | Ottobock
9 Nm/6.6 ft.lbf./80 lbf.in.
7 Nm/5.2 ft.lbf./62 lbf.in.