5. Cryotips can be steam sterilized by Gravity or Pre-vacuum using the following procedures:
Set-up
Pre-Exposure
(Conditioning)
Exposure
(Steam Injection)
Post-Exposure
(Exhaust / Dry)
6. Attach a new disposable plastic shield to the cryotip, using an aseptic technique. Ensure that the tabs
on the shields are correctly fitted into the grooves on the metal tip.
7. The cryotip is now ready to use.
WHERE TO REFILL GAS
TO ENSURE OPTIMAL PERFORMANCE, ONLY MEDICAL GRADE CO
CONTAMINANTS SHOULD BE USED.
Medical grade CO
of these headings: Welding Supplies, Gas-Medical, Carbonic Gas, Soda Dispensing Equipment and
Supplies. Many of these suppliers deliver to the office.
WARRANTY
Wallach Surgical guarantees the Wallach LL-CO
of five (5) years against defects in material and workmanship.
SERVICE AND REPAIR
In case of a malfunction, immediately turn the gas cylinder master valve completely off. Detach
instrument from gas cylinder. Call the factory at (203) 799-2000 and ask the Service Department
for assistance.
This device does not have any user serviceable parts.
If a repair is needed, carefully sanitize the unit and package it in a protective carton. Include a note
stating the problem and ask for a repair estimate. Do not attempt in-house repairs as they will void
your warranty.
37295 • Rev. B • 2/12
™ Cryosurgical System • Directions for Use (English)
LL-CO
2
Stage / Process
Cycle Profile
Chamber Purge
Sterilization Temperature
Sterilization Time
Dry Time
gas is available throughout the world. See your local telephone listing under any
2
Parameter Set-Points
Gravity
1 minute
132 °C (270 °F)
15 minutes
15 minutes
™
Freezer Cryosurgical System and Tips for a period
2
5
Pre-vacuum
1 minute
132 °C (270 °F)
4 minutes
20 minutes
GAS FREE OF
2
Wallach Surgical Devices