DC Current
Ranges
Resolution
Accuracy
Burden voltage
Input protection
AC Current true rms (45Hz - 1kHz)
Ranges
Resolution
Accuracy
Crest Factor
Burden voltage
Input protection
400μA, 4000μA, 40mA, 300mA, 10A
0.1μA
±(1.0 % rdg + 1 dgt) on 400μA to 300mA ranges
±(2.0 % rdg + 3 dgts) on 10A range.
400 μA Range:
4 mA Range:
40 mA Range:
300 mA:
10A:
0.315A/1000V fast blow ceramic fuse
6.3×32mm on μA/mA input 10A/1000V fast
blow ceramic fuse 10×38mm on 10A input
10A input: 10 A for 5 minutes maximum
followed by a 10 minute cooling period
400μA, 4000μA, 40mA, 300mA, 10A
0.1μA
±(1.5 % rdg + 8 dgts) on 400μA to 300mA ranges
±(2.5 % rdg + 10 dgts) on 10A range
≤ 3
See DC Current
0.315A/1000V fast blow ceramic fuse
6.3×32mm on μA/mA input 10A/1000V fast
blow ceramic fuse 10×38mm on 10A input
10A input: 10 A for 4 minutes maximum
followed by a 12 minute cooling period
11
1 mV/ 1 μA
500 mV/ 1 mA
10 mV/ 1 mA
8 mV/ 1 mA
40 mV/ 1 A