Gas Purge Valve Installation; Pump Used With Corrosive Gases - Varian Turbo-V 2000HT Manual De Instrucciones

Ocultar thumbs Ver también para Turbo-V 2000HT:
Tabla de contenido

Publicidad

GAS PURGE VALVE INSTALLATION

A gas purge valve is available to protect the pump
bearings against particulate and corrosive gases
that could move into the pump.
To install the gas purge valve it is necessary to
unscrew the purge port cover as shown in the fol-
lowing figure,
and then screw the gas purge valve (with a torque
of 2.5 Nm) as shown in the following figure.

PUMP USED WITH CORROSIVE GASES

To prevent damage to the bearings, an inert gas
must flow into the pump body around the upper
bearing towards the forevacuum line.
To supply the inert purge gas (e. g. nitrogen) to
the pump through the purge port, connect a gas
purge valve between the pressure regulator and
the pump (see the following figure). Adjust the
pressure regulator in order to obtain the necessary
gas purge flow rate according to the application.
To prevent bearing damage, Varian suggests the
following purge gas flow rates: 10 sccm (0.17
mbar l/s) for a mild purge process, or 20 sccm
(0.34 mbar l/s) for an heavy process. Please con-
tact Varian for specific applications.
The purge gas throughput with the recommended
forepump of 15 m
a high vacuum pressure in the 10
The recommended gas flow maintains a pressure
into the pump body higher than the forevacuum
pressure.
31
TECHNICAL INFORMATION
!
CAUTION
3
/h (11 CFM) allows to achieve
-9
mbar range.
87-900-914-01(F)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969-9084 serie969-9085 serie