When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.
Note: When a printer is connected the print menu opens and the result can be
printed.

View the calibration data

Data from the most recent calibration can be shown.
1. From the main menu use the
Confirm.
2. Select Display data.
3. Select Calibration data and confirm with
is shown.
• pH-the slope and offset values are shown alternating with the
deviation (in %) and calibration temperature.
• ORP-the measured mV value and calibration temperature are
shown.
• Conductivity-the cell constant and calibration temperature for each
standard are shown.

Set the calibration reminder

pH: The calibration reminder can be set between 0 to 23 hours or
1-7 days (default 1 day). EC: The calibration reminder can be set
between 0 to 99 days (default 15 days). The display shows the
remaining time to the new calibration.
Note: When 0 days is selected, the calibration reminder is turned off.
1. From the main menu use the
Confirm.
2. Use the
to enter the calibration menu.
3. Use the
or
to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the
change a value. Confirm.
Push
to start the calibration.
or
to select DATA LOGGER.
. The last calibration data
or
to select CALIBRATION.
or

Sample measurements

Each probe has specific preparation steps and procedures for taking
sample measurements.
1. From the main menu use the
MEASURE. Confirm.
2. Use the
to change the following settings. Confirm every entry.
Option
Description
Resolution
Select the resolution: 1, 0.1, 0.01 (default) or 0.001
Measure
Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.
Display mV
Display mV—select YES or NO to show mV.
Limits
Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.
Isopotential Isopotential—change the Isopotential pH value in Data
introduction. Select Calculate to calibrate the probe again.
TC
TC—select Linear or Natural waters. Linear: enter a value in
%/°C (default 2.00%/Temperature). Natural waters: Non-linear
for natural waters, according to EN27888
Tref
Reference temperature—select between 20 or 25 °C or Other
temperature.
3. Push
to start the measurement.
Note: If the measurement is not stabilizing after 120 seconds, the meter turns
to
automatically into the continuous measurement mode.
or
and and to select