The next calibration solution is shown.
9. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the probe
measuring chamber.
10. Push
to measure the third calibration solution.
When the calibration is good, the display briefly shows Calibration OK
and then returns to the main menu.
Note: When a printer is connected the print menu opens and the result can be
printed.
View the calibration data
Data from the most recent calibration can be shown.
1. From the main menu use the
Confirm.
2. Select Display data.
3. Select Calibration data and confirm with
is shown.
• pH-the slope and offset values are shown alternating with the
deviation (in %) and calibration temperature.
• Conductivity-the cell constant and calibration temperature for each
standard are shown.
Set the calibration reminder
The calibration reminder can be set between 0 to 23 hours or 1-7 days
(default 0 days). The display shows the remaining time to the new
calibration.
Note: When 0 days is selected, the calibration reminder is turned off.
1. From the main menu use the
2. Use the
to enter the calibration menu.
3. Use the
or
to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the
change a value. Confirm.
Push
to start the calibration.
or
to select DATA LOGGER.
. The last calibration data
or
to select CALIBRATION. Confirm.
or

Sample measurements

Each probe has specific preparation steps and procedures for taking
sample measurements.
1. From the main menu use the
Confirm.
2. Use the
to change the following settings. Confirm every entry.
Option
Description
TC
TC—select Linear or Natural waters. Linear: enter a value in %/
°C (default 2.00%/Temperature). Natural waters: Non-linear for
natural waters, according to EN27888
Tref
Reference temperature—select between 20 or 25 °C or Other
temperature.
TDS Factor Total dissolved solids (TDS) factor—enter TDS Factor
Measure
Stability—select By stability Criterion: ± 1 digit in 6 s. In
continuous—enter the time interval for the In continuous Acquis.
interval. By time—enter the time interval for data storage or
printing data.
Display Ω
Display Resistivity—select YES to show simultaneously
conductivity and resistivity.
3. Push
to start the measurement.
Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.

Advanced operations

Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1. From the main menu use the
to
2. Use the
or
to select Sample ID and confirm.
or
and and to select MEASURE.
or
to select SYSTEM. Confirm.
English 9