Option
SET STANDARD LEVEL
SET CLEANER LEVEL
RESET REAGENT
LEVELS
RESET STANDARD
LEVELS
RESET CLEANER LEVELS Sets the volume of the cleaning solution in the cleaning solution bottle to 100%
PRIME REAGENTS

Grab sample options

The grab sample menu allows the user to analyze a specific sample or standard. Two options are
available:
• GRAB SAMPLE IN: This option is used to analyze an external sample or standard.
• GRAB SAMPLE OUT: This option is used to extract a sample directly from a sample line for
external analysis.
Measure a grab sample or standard
Use the grab sample funnel to analyze grab samples taken from other areas in the water treatment
process or to measure a standard for a calibration verification. A minimum of 300 mL of sample is
necessary for this procedure.
1. Push menu and select GRAB SAMPLE>GRAB SAMPLE IN.
2. Complete the instructions on the display.
Get a grab sample from the analyzer
Complete the instructions on the screen to dispense a grab sample from one of the sample sources
for an external analysis. Make sure to use a container that can hold 300 mL or more.
After the sample is dispensed, the analyzer immediately measures the sample source. The value of
the sample source is shown on the display.
1. Push menu and select GRAB SAMPLE>GRAB SAMPLE OUT.
2. Complete the instructions on the display.

Set up the system

The configuration settings can be changed in the SETUP SYSTEM menu or in the analyzer setup.
Refer to
Start the analyzer setup
1. Push menu and select SETUP SYSTEM.
2. Select an option.
Option
SETUP ANALYZER
MEAS UNITS
10 English
Description
Sets the volume of a specific standard solution in the standard bottle to a
calculated approximate value. Range: 1–100%.
Sets the volume of the cleaning solution in the cleaning solution bottle to an
estimated value. Range: 1–100%.
Sets the volume of reagent in the reagent bottles to 100% full.
Sets the volume of standard solution in the standard bottle to 100% full.
full.
Starts the flow of all reagents through the tube and valve system.
on page 8.
Description
Sets the analyzer to off and then to on again.
Changes the measurement units that are shown on the display and in the
data log. Options: ppm (default), mg/L, ppb, µg/L.