Do a standard calibration adjust
1. Complete the test procedure for the range to calibrate. For the
sample, use the standard solution concentration given in the test
procedure documentation.
Note: If a standard solution concentration is not given in the test procedure
documentation, a different known standard can be used.
2. When the test procedure is completed, push and hold
3 seconds.
3. Push
until "SCA" shows, then push
The display shows the standard calibration adjust value.
4. If a different known standard is used, enter the value of the
standard:
a. Push
b. Enter the value of the standard.
Push the
go to the next digit. Push
5. Push
to add the standard calibration adjust value to the factory
calibration curve.
Set the standard calibration adjust to off
To use the factory default calibration again, set standard calibration
adjust (SCA) to off.
1. Push and hold
2. Push
until "SCA" shows, then push
3. Push
until "OFF" shows, then push
Note: To set the SCA function to on again, do a standard calibration adjust.
User-entered calibration curve
This instrument accepts a user-prepared calibration curve. The
calibration curve can be from 0 to 2.5 absorbance. Make sure that the
calibration curve includes standard values that are less and more than
the range of interest.
The instrument range will be the same as the calibration range. For
example, when the standards that are used are 1.00, 2.00 and 4.00.
The instrument range is 1.00 to 4.00.
16 English
until "Edit" shows, then push
or
to change the number that flashes. Push
for 3 seconds to enter menu mode.
.
.
to go to the previous digit.
.
.
for
to