Set temp.
Set MTC temperature
When you are working without a temperature sensor, enter the temperature of the sample in this menu
(-5 °C...105 °C). The meter shows the temperature-compensated value in the measurement display.
Set meas.
EP format
With this menu you can choose between one of three different types of endpoint formats:
"Auto"
Automatic endpoint
is a special algorithm that determines the end of an individual reading, depending
on the behavior of the sensor used.
Stability criterion for conductivity measurements
The measured conductivity of the sample may not deviate by more than 0.4% from the measured average
conductivity in 6 seconds.
"Manual"
Manual endpoint
means the meter never endpoints the reading unless the user manually presses Read.
"Timed"
With the timed endpoint
Note
Every measurement can be endpointed manually by pressing the Read key. The meter then shows
The table below shows how the entpoint format is displayed in the course of the measurement.
Preselected format
Auto endpoint
Manual endpoint
Timed endpoint
The actual enpoint format (last column) not the preselected is stored with the data.