Technical Specifications/Certifications; Technical Specifications Of The Altair - MSA ALTAIR Manual De Operación

Ocultar thumbs Ver también para ALTAIR:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 99

TECHNICAL SPECIFICATIONS/CERTIFICATIONS

5.
Technical specifications/Certifications
5.1.

Technical specifications of the ALTAIR

Weight
Dimensions
Alarms
Volume of acoustic alarm 95 dB typical
Displays
Battery type
Battery life
Sensor
Measuring range
Factory alarm setpoints
Temperature range
Humidity range
Dust and spray
protection
Function tested gases
Warranty
*)
When placing an order, you can request different alarm setpoints. The settings can be changed before
commissioning with the help of the test button and thereafter, with the FiveStar
28
125 g (instrument with battery and clip)
86 x 51 x 50 mm (L x B x T) – with fastening clip
Two super bright LEDs with 320° viewing angle and
a loud acoustic alarm
Large display of measured values
3.6 V dry batteries ½ AA (Lithium),
not replaceable
Approx. 24 months under normal conditions
Electrochemical
H
S
2
0-100 ppm
*)
LOW ALARM
CO
H
S
2
O
2
Operation
Storage
Vibration alarm
10 - 95% rel. humidity, non-condensing
IP 67
S, O
CO, H
2
2
2 years after activation or
18 hours alarm time on commissioning within 6
months after manufacture
ALTAIR Operating Manual
CO
0-500 ppm
HIGH ALARM
25 ppm
10 ppm
19.5%
:
- 20°C to +50°C
:
0°C to +40°C
:
to 0°C
®
Link
MSA
O
2
0-25 % Vol.
100 ppm
15 ppm
23.0%
TM
software.
GB

Publicidad

Tabla de contenido
loading