11. Slide sensor into a mounting frame back end first aligning the angled edge of the sensor
12. Tighten the sensor clamping screws to hold the sensor firmly in place.
13. Observing the upstream and downstream orientation, attach the UP and DN cables to the
3.4.3

Final Setup

1. At [Install Sensor] menu, scroll down to [Install Complete]. Press the <Right Arrow> and
2. Observe the Measured Vs window and verify a correct sound velocity measurement (if
3. Press the <Down Arrow> to accept sound velocity value.
4. The meter is now ready to report flow. Press the <MENU> key twice to display flow.
FUP1010 IP67 Portable Quick Start
Operating Instructions, 01/2013, CQO:QSG003 Revision 05
with the angled edge of the mounting frame. Keep sensor from making contact with the
pipe until it butts up against the mounting frame stop. Push sensor down to mate with
pipe.
for the other sensor
sensors and make snug. Attach the other ends to the UP and DN terminals of the (see
figure below).
Flow Sensor Cable Connectors
To Channel 2 Flow Sensor Set
Figure 3-9
Connecting Sensors to Flow Meter
select [Install]. Press <ENTER>. Flow meter will go through drives.
known).
3.4 Sensor Installation
Channel 1 - Upstream Sensor
Channel 1 - Downstream Sensor
Commissioning
Repeat procedure
29