3.5

Starting and stopping the centrifuge

3.5.1 Starting the centrifuge

To start the centrifuge, make sure that the lid is closed and then press the key "Start" (9), (refer to figure 22). With
the key "Start" (9) you can start runs with manually pre-selected parameters. When the respective pre-selected
running time has ended then the centrifuge will stop automatically or you can interrupt each run with the key
"Stop" (10).
Figure. 22

3.5.2 The "STOP" key

The run can be interrupted at any time by pressing the "Stop" key (10) (See figure 23). After pressing the key the
centrifuge decelerates with the respective pre-selected intensity until it reaches standstill.
Figure. 23
3.6

Imbalance detection

In case of the rotor not being equally loaded, the drive will turn off during acceleration. The rotor will decelerate to a
standstill.
When in the display "Time" (A-3) the word "error" (M11) together with the number "01" appear, the weight
difference of the samples is too large. Weigh the samples more exactly and start again.
Load the rotor as described in chapter 3.1.2 and 3.1.3.