Specification
Temperature range:
Power supply:
Environmental protection:
Altitude:
Fuses:
Safety:
E.M.C:
Dimensions:
Weight:
Cleaning:
*Relates to transient overvoltage likely to be found in fixed installation wiring
18
Operating: -20 °C to +50 °C (Errors double outside range 0 to 30 °C)
Storage: -25 °C to +65 °C
Two 12 V, 2 Ah Lead-acid rechargeable batteries. Recharge time: 16 hours
Battery life: typically 8 hours continuous testing
IP54 (with the charging recess cover securely closed)
2000 m max. to operate within specification
FS1: 100 mA (T), 250 V IEC 127/1
FS2: 100 mA (F), 250 V IEC 127/1
Mains Power cord fused plug (when applicable): 3 A fuse to BS 1362
The instrument meet the requirements for double insulation to IEC 1010-1 (1995) EN
61010-1 (1995) to installation Category III*, 300 Volts phase to earth (ground) and 500
Volts phase to phase
In accordance with IEC 61326 including amendment No.1
327 mm x 316 mm x 196 mm (+ 60 mm pouch)
6 kg
Wipe disconnected instrument with a clean cloth dampened with soapy water or Isopropyl
Alcohol (IPA).