Sensor change function
With the menu function Change Sensor, a sensor can be
changed during operation without triggering a fault signal at
the control unit. It also ensures that all sensor data in the
microprocessor can be saved in the sensor memory
(EEPROM) before disconnecting the sensor.
Replacing the sensor
A sensor can be replaced at any time.
1. Select Settings > Sensor > Change Sensor and confirm.
 The maintenance signal is generated on the 4-20mA
interface.
 The display shows Please remove the sensor.
 The Maintenance symbol is displayed.
2. Replace the old sensor with a new sensor:
a. Loosen adjusting screw.
b. Unscrew the bayonet ring from the housing.
c. Remove and discard the cover plate or old sensor
respectively.
d. Remove the sensor from packaging and ensure that
the sensor is a DrägerSensor.
e. Insert the new sensor into the sensor opening with the
Dräger logo facing up.
f. Reinstall bayonet ring over the sensor, and turn
clockwise until sensor is locked in place.
g. Tighten adjusting screw.
When the new sensor is installed, the display shows:
"Loading data, please wait".
When the sensor data are loaded, the display shows:
"Data loaded".
3. Select Back to menu and confirm.
 The maintenance signal on the 4-20mA interface
remains pending until the sensor is warmed-up.
4. Check calibration. If necessary, calibrate the instrument
(see "Calibration", page 16).

10.3.2 PointGard 2200 CAT

Replacing the sensor
This point is only valid for the PointGard 2200 CAT without
remote application.
1. If necessary, set the maintenance current for the analog
interface.
2. Open the instrument.
3. Unplug the sensor connector from the socket.
4. Unscrew the sensor and pull it out of the port.
5. Route the cables of the new sensor through the port for
the sensor.
6. Screw the new sensor in the port for the sensor with
correct torque (194 +/- 17 Lbs. In.22 +/- 2 Nm / 194 +/- 17
Lbs. In.22 +/- 2 Nm).
7. Plug the sensor connector to the socket.
8. Close the instrument.
9. Commission the instrument.
Instructions for use
|
Dräger PointGard 2000 Series
10. Check calibration. If necessary, calibrate the instrument
(see "Calibration", page 16).
Replacing the remote sensor
1. If necessary, set the maintenance current for the analog
interface.
2. Proceed as described in the manual of the sensing head.
3. Check calibration. If necessary, calibrate the instrument
(see "Calibration", page 16).

10.3.3 PointGard 2700 IR Remote

Replacing the remote sensor
1. If necessary, set the maintenance current for the analog
interface.
2. Proceed as described in the manual of the sensing head.
3. Check calibration. If necessary, calibrate the instrument
(see "Calibration", page 16).

10.4 Performing display test

1. Select Settings > Instrument > Display > Displaytest
and confirm.
2. Select Enable or Disable and confirm.
When enabled, the display is shown inverted and all the
LEDs are lit.
11

Instrument settings

For further information and configurations, consult the
technical manual 9300148.

11.1 Setting passwords

1. Select Settings > Instrument > Passwords the desired
password and confirm.
Calibration
Access to zero and span calibration
PWD
Settings PWD Access to all configuration parameters
2. Select the line for editing the password and confirm.
3. Set the password and confirm.
4. Select Confirm and confirm with [OK].

11.2 Setting date and time

1. Select Settings > Instrument > Date and time and
confirm.
2. Select the line for editing the date or time and confirm.
3. Select Confirm and confirm with [OK].

11.3 Setting language

1. Select Settings > Instrument > Language and confirm.