Specifications; Classification Of Equipment; Operation Principle - NSK VarioSurg 3 Manual De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
13

Specifications

<Control Unit>
Model
NE288
Rated voltage
AC120V 50/60Hz
AC230V 50/60Hz
Drive frequency
28 - 32kHz
Maximum output
25W
Power input
54VA
Dimension
W265 x D220 x H103mm
Weight
3kg
<Foot Control>
Model
FC-78
Cord Length
2m
Dimension
W268 x D230 x H103mm
Weight
1.4kg
Use environment
Transportation and Storage
environment
* No liquid freezing allowed
* When operated outside the above range, there is a risk of failure.
14

Classification of Equipment

Type of protection against electric shock
- Class I equipment
Degree of protection against electric shock:
- Type BF applied part
Method of sterilization or disinfection recommended by the manufacturer:
- See"9-6 Sterilization"
Degree of protection against ingress of water as detailed in the current edition of IEC 60529:
- Foot Control: IPX8 (Protected against the effects of continuous immersion in water)
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide:
- Equipment NOT suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide. (Without Foot Control)
Foot Control: Category AP Equipment
Mode of operation:
- Continuous operation
15

Operation Principle

A sinusoidal electrical signal, at ultrasonic frequency (f>20kHz), is delivered by the generator. This signal is applied to the
'piezoelectric ceramic' located inside the transducer. Piezoelectric ceramic converts this signal into mechanical vibrations.
These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations are propagated
towards the distal end of the transducer. The "TIP" insert, which is attached at the distal end of the transducer, vibrates at
ultrasonic frequencies and makes it possible to achieve the aimed purpose.
Temperature
0 - 40°C
-10 - 50°C
(Applied part: Tip, Handpiece)
<Handpiece>
Model
Cord Length
Type of oscillation
Optic
Consumption current(LED) 0.17A (3.5V)
Dimension
Weight
Humidity
Atmospheric pressure
30 - 75%
10 - 85%
500 - 1060hPa
20
VS3-LED-HPSC
2m
Piezo type
White LED
ø20×129mm (except cord)
70g (without cord)
-

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido