Do not handle the power cord with wet hands. Wet hand contact with electricity may result in an electric shock.
Keep away from explosive substances and flammable materials.
If the product overheats or smells of burning, immediately turn off the power and disconnect the main power plug.
Contact your Authorised NSK Dealer.
TO PREVENT ELECTRIC SHOCK, use a main electrical outlet that is protective earthed.
Do not use on the following patients.
-Those with medical complications or allergies
-Those who have pre existing conditions
(Eg Cardiac, Pulmonary, Renal disturbance or High blood pressure)
-Those who are pregnant or lactating
-Patients with cardiac pacemakers and infants
Be careful not to get water or liquid disinfectant on the control unit. This could cause short circuits and lead to fire and/
or electric shock.
Repeatedly turning the Main Power Switch ON and OFF may blow a fuse.
When installing the product, provide space of approximately 10cm around the product for easy access to the inlet and
the Power Cord.
The Irrigation Tubes, included in the product package, are sterile. When using these products, follow the instructions
below.
- The irrigation tube is a single-use item. Do not reuse nor sterillize as product breakage or infection could occur.
- Check before use that the sterile package is NOT open nor damaged. Do not use products with open or damaged
packages as sterility will be negated and infection could occur.
- Observe the use-by date written on the sterile package label. Do not use expired products as sterility may be
negated.
Do not point the handpiece LED light directly to the eyes of the patient, nor the operator, as it may be harmful to eyes.
To be confident in using this product, please read this Operation Manual before use. We suggest that to fully
understand the product functions and file for future reference.
When operating the product always consider the safety of the patient.
The end user shall be responsible for any judgment that relates to the application of this product to a patient.
This product does not consider patient's age (except infants), gender, weight or nationality.
This product does not consider operator's age (mature person), height, weight, gender, or nationality.
Users are responsible for the operational control, maintenance and continual inspection of this product.
This device is for indoor use only.
Keep the Control Unit on a level surface.
Do not attempt to disassemble the product or tamper with the mechanism except as recommend by NSK in this
Operation Manual.
Do not allow any impact on to the handpiece. Do not drop the handpiece.
Operators and all others in the area must wear eye protection and a mask when operating this handpiece.
Should the product function abnormally, cease operation immediately and contact your Authorized NSK Dealer.
If the Irrigation Pump gets wet, wipe it dry. If it remains wet, the roller in the pump may slip and the pump may not
operate properly.
Do not bend or fold the Irrigation Tube while the Irrigation Pump is operating as the Tube may be damaged or may
detach.
If abnormality is detected in irrigation, it may be due to wear of the Irrigation Tube or leaking of saline from the Tube. In
such cases, replace the Irrigation Tube.
Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
The following products are delivered in a non-sterile condition and must be autoclaved prior to use. (Handpiece, Tip, Tip
Wrench, Tip Holder, Tube Holder)
Perform regular function and maintenance checks.
If the product is not used for a long period check it is functioning correctly before using on a patient.
To avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown during surgery.
This product may be affected by an electric scalpel. Turn OFF the Power Switch when an electric scalpel is used.
WARNING
CAUTION
2