Electromagnetic Compatibility (Emc) - Walgreens Well Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ELECTROMAGNETIC COMPATIBILITY (EMC)

• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used in such
environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in such
environments:
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
• Recommended separation distances between portable and mobile RF communication equipment and the device.
The device is intended for use in an electromagnetic environment where radiated RF disturbances are under control. User
can help prevent electromagnetic interference by keeping the device at a minimum distance from portable and mobile RF
communications equipment (transmitters). Below table details the maximum output power of transmitter:
Rated maximum
output power of
transmitter W
0.01
0.1
1
10
100
Compliance
Electromagnetic environment – guidance
Group 1
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not likely to
Class B
cause any interference in nearby electronic equipment.
Not
The device is suitable for use in all establishments, including
Applicable
domestic establishments, and those directly connected to
the public low-voltage power supply network that supplies
Not
buildings used for domestic purposes.
Applicable
Compliance level
± 8 kV contact
± 15 kV air
30 A/m
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
Not Applicable
d = 1.2
NA
NA
NA
NA
NA
Electromagnetic environment–
guidance
The relative humidity should be at
least 5 %.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Recommended separation distance
I
r = --------- (m)
188
where I is the current in amperes in a
power bus or an appliance wire and r is
the recommended separation distance
between your device and the power bus
or appliance wire, in meters (m).
800 MHz to 2.5 GHz
d = 2.3
0.12
0.38
1.2
3.8
12
31
0.23
0.73
2.3
7.3
23

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido