Siemens SITRANS Probe LU Guia De Inicio Rapido página 12

Ocultar thumbs Ver también para SITRANS Probe LU:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59
Quick Setup: steps 1 to 10
Default values (factory settings) are marked by an asterisk (*) in the tables.
Note:
1. Select language (P010: Language)
Numeric/None
0 *
English
1
German
Values
2
French
3
Spanish
4
2. Set P001: Operation (measurement mode)
Notes:
• Setting P001 resets Span (P007), unless Span has previously been set to a different value.
• Changing P001 may reset Output Function (P201): this applies to HART only.
Level returns distance to material level referenced from Empty (process empty level). The
1 *
reading is returned in volumetric units if parameters 050 to 055 are set to enable this.
Values
2
Space returns distance to material level referenced from Span (process full level).
3
Distance returns distance to material level from reference point (transducer face).
mA Output with Level, Space, and Distance operation
Level
(P001 = 1)
transducer
face
Span
P007
P006
Empty
Page EN-10
SITRANS Probe LU (HART) – QUICK START MANUAL
ENGLISH
P000
LOCK
P001
OPERATION
P003
MEAS RESP
P004
ANTENNA
P005
UNITS
P006
EMPTY
P007
SPAN
P010
LANGUAGE
Space
(P001 = 2)
(20 mA)
Span
100%
Level
Empty
0%
(4 mA)
DEUTSCH
FRANÇAIS
VERRIEGELG
VERROUIL
BETRIEB
FONCTIONMT
REAKTIONSZ
TEMPS REP .
ANTENNE
ANTENNE
EINHEIT
UNITES
MESSBER.
VIDE
MESSSPANNE
PLAGE
SPRACHE
LANGUE
Distance
(P001 = 3)
(4 mA)
0%
Space
100%
(20 mA)
ESPAÑOL
BLOQUEO
FUNCIONAM.
TIEMPO R.
ANTENA
UNIDADES
VACIO
RANGO
IDIOMA
0%
(4 mA)
Span
Distance
Empty
100%
(20 mA)
A5E32168031

Publicidad

Tabla de contenido
loading

Tabla de contenido