NOTE: Scale can only be accurately calibrated using standard calibrated scale weights. Using
non-calibrated weights may result in inaccurate scale calibration. Nothing should be on the scale and
scale should be level.
ACTIVATE WEIGHT CALIBRATION MODE
• Switch the mini jumper SWA1 to "ADJ" position. Adjust is down; Lock is up. The switch is located under
0
a panel on the right side of the keypad housing.
• Turn on the scale. As scale counts down to zero, press and hold the ZERO key until the software
version (ex:
0
• Release ZERO key.
• Press ENTER key. Display shows
• Press ENTER key. Display shows
• Press UNITS or Kg key to set scale to lb or kg as needed.
UNITS
Kg
• Use the UP and ENTER keys to preset the weight being used to calibrate the scale. The UP key sets the
number and the ENTER key moves right one digit.
Example: To set a weight of 3.0 kg, you would press the ENTER key three times to move right three
spaces and the UP key three times to enter the number 3.
• Place your calibrated weight on the scale.
• Press ENTER key. Display shows
• Scale displays weight and beeps.
• Scale is calibrated, it will now return to weighing mode automatically.
• Switch the mini jumper SWA1 back to "LOCK."
π
CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA
PAGE 4 OF 12

CALIBRATION INSTRUCTIONS

02018
) appears on the display.
F 0
appears in display.
ZEro
0---,-Ckg
WAIt
1-800-295-5510
.
.
.
uline.com
0120 IH-670

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

H-747H-794H-4593

Tabla de contenido