LEVELING THE SCALE
Unit has an illuminated level indicator as a reminder
that the scale should be leveled for accurate weighing.
There is a level bubble in a small round window on
the front of the scale. To level, adjust the feet at each
corner until the bubble is centered in the circle. Be sure
the equipment is level each time its location is changed.
(See Figure 5)
Figure 5
Function
Tare
Mode
Menu
Back
Exit
CONNECTING POWER
AC ADAPTER INSTALLATION
NOTE: AC power is used to power the scale
when battery power is not needed.
Connect the AC adapter (included) to the AC adapter
input jack at the rear of the scale, and connect the AC
plug to an electrical outlet. (See Figure 5)
BATTERY INSTALLATION
Install the four AA batteries with polarity as shown in the
battery compartment. (See Figure 6)
NOTE: After powering on the unit, it is
recommended to let it warm for at least five
minutes before using.
Figure 6
AC Adapter
Input Jack
Interface
Connection
PAGE 3 OF 53
INSTALLATION CONTINUED
Level Bubble
Level
Not Level
Battery
Compartment
Cover

INITIAL CALIBRATION

When the scale is first installed, and when it is moved
to another location, it must be calibrated to ensure
accurate weighing results.
Scale must be calibrated by using the external
calibration weights.
Have the appropriate calibration masses available
before beginning calibration. Refer to
Calibration Operation section on page 6 for
masses and calibration procedure.