Mectron PIEZOSURGERY Manual Del Usuario página 3

Limpieza y esterilización de los insertos reusables
Ocultar thumbs Ver también para PIEZOSURGERY:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

01 Û INTRODUCTION
Û TABLE OF CONTENTS
01 Û INTRODUCTION
EN
02 Û CLEANING AND STERILIZATION
02.1 Pre-cleaning with enzymatic detergent . . . . . . . . . . . . . . . . . . . . . . . . . . 3
02.2 Ultrasonic bath or Automated cleaning and thermodisinfection . . . 5
02.3 Inspection and drying . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
02.4 Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
01 Û INTRODUCTION
This manual describes the cleaning and
sterilization procedures to be applied on
reusable bone surgery inserts produced by
Mectron.
Carefully read this manual before proceeding
with the cleaning and sterilization operations,
and always keep it within reach.
02 Û CLEANING AND STERILIZATION
The operator must perform a cleaning cycle
with enzymatic detergent (PRE-CLEANING)
as first step. After that it can be decided to
perform the:
- Ultrasonic bath
- Automated cleaning and thermodisinfection.
The procedures applied on PIEZOSURGERY®
accessories have been validated by an
independent institute.
WARNING: First use: The inserts are
delivered in NON STERILE condition and
must be prepared prior the use by applying
the procedures described in this manual.
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
IMPORTANT: to prevent harm to persons
or damages to objects, read all the "Safety
precautions"present in the manual with
special care. Depending on their degree
of seriousness, the safety precautions are
classified with the following indications:
WARNING
(always refers to personal injury)
CAUTION
(refers to possible damage to property)
Every use: Once used, each reusable insert
must be thoroughly reprocessed prior
to reuse, according to the procedures
described in this manual.
WARNING: Cleaning processes must
immediately be carried out after each use.
Do not allow contaminated inserts to dry
prior to reprocessing.
Limitations on reprocessing: Repeated
processing has minimal effect on the
instruments examined in this manual. End
of life is normally determined by wear and
damage due to use.