7

Sensor Calibration

h
Note: For best process calibration results observe the following points. Take grab sample as
close as possible to the measurement point of the sensor. Measure the sample at process
temperature.
For a detailed description of the "Process", "1-Point" and "2-Point" calibration method see
operating instructions of the M300 transmitter. It is also possible to connect pre-calibrated
sensors, calibrated via iSense software.
As soon as the calibration is in progress no other calibration can be started.
Sensor Calibration Menu c
After every successful calibration different options are available. If "Adjust", "SaveCal" or "Cal-
ibrate" is chosen, the message "Calibration saved successfully! Reinstall sensor" is displayed.
Press "Done".
Option
Analog sensors:
Save Cal
ISM sensors:
Adjust
Calibrate
Cancel
Analog sensors
Calibration values are stored in the
transmitter and used for the measure-
ment. Additionally, the calibration val-
ues are stored in the calibration data.
The function "Calibrate" is not applica-
ble for analog sensors.
Calibration values are discarded.
ISM (digital) sensors
Calibration values are stored in the sen-
sor and used for the measurement. Ad-
ditionally, the calibration values are
stored in the calibration history.
Calibration values are stored in the cali-
bration history for documentation, but
not be used for the measurement. The
calibration values from the last valid
adjustment are further used for the mea-
surement.
Calibration values are discarded.