7

Sensor Calibration

h
Note: For best process calibration results observe the following points. Take grab sample as
close as possible to the measurement point of the sensor. Measure the sample at process
temperature.
For a detailed description of the "Process", "1-Point" and "2-Point" calibration method see
Operation Manual of the M200 transmitter. Via iSense software you can calibrate the sensor
with the "1-Point" or "2-Point" calibrate method. See Operation Manual of the iSense software.
As soon as the calibration is in progress no other calibration can be started.
Sensor Calibration Menu
After every successful calibration the options "Yes" or "No" are available. After selection one of
these options the message "Re-install sensor and Press [ENTER]" appears.
After pressing [ENTER] the M200 returns to the Measurement mode.
• Yes
Calibration values are stored in the sensor and used for the measurement.
• No
Calibration values are discarded.