Sensor Calibration

h
Note: For best process calibration results observe the following points. Take grab sample as
close as possible to the measurement point of the sensor. Measure the sample at process
temperature.
For a detailed description of the "Process", "1-Point" and "2-Point" calibration method see Op-
eration Manual of the M800 transmitter. Via iSense software you can calibrate the sensor with
the "1-Point" or "2-Point" calibrate method. See Operation Manual of the iSense software.
As soon as the calibration is in progress no other calibration can be started.
Sensor Calibration Menu c
After every successful calibration different options are available. If "Adjust", "SaveCal" or "Cali-
brate" is chosen, the message "Calibration saved successfully! Reinstall sensor" is displayed.
Press "Done".
Option
Analog sensors:
Save Cal
ISM sensors:
Adjust
Calibrate
Cancel
Analog sensors
(1-channel version only)
Calibration values are stored in the
transmitter and used for the measure-
ment. Additionally, the calibration values
are stored in the calibration data.
The function "Calibrate" is not applicable
for analog sensors.
Calibration values are discarded.
ISM (digital) sensors
Calibration values are stored in the sen-
sor and used for the measurement. Ad-
ditionally, the calibration values are
stored in the calibration history.
Calibration values are stored in the cali-
bration history for documentation, but
not be used for the measurement. The
calibration values from the last valid
adjustment are further used for the mea-
surement.
Calibration values are discarded.