Sensor Calibration

(PATH: Device setup / Sensor Calibration)
Sensor Calibration Menu
12
h
Note: For best process calibration results observe the following points. Take
grab sample as close as possible to the measurement point of the sensor.
Measure the sample at process temperature.
h
Note: Via configuration tool, asset-management-tool or HART handheld ter-
minal you can calibrate the sensor with the "Process calibration" method.
For other calibration methods use iSense software.
h
Note: As soon the calibration is in progress no other calibration can be
started.
For the calibration methods see "Menu Structure" chapter on Page 9.
After every successful calibration the following options are available:
Adjust
Calibration values are adopted and used for the measurement. Additionally, the
calibration values are stored in the calibration history.
Calibrate
Calibration values are stored in the calibration history for documentation, but not be
used for the measurement. The calibration values from the last valid adjustment are
further used for the measurement.
Abort
Calibration values are discarded.