17. EMC Technical results

Test
Emission (IEC 60601-1-2 section 7)
Conducted
emission
Freq range 150
kHz – 30 MHz
Radiated emission
Freq range 30 –
1000 MHz
Harmonic current
emission test
Voltage changes.
Voltage
fluctuations and
Flicker test
Immunity (IEC 60601-1-2 section 8)
Immunity from
Electrostatic
discharge (ESD)
Immunity from
radiated
electromagnetic
fields
Immunity from
Proximity field
from wireless
communications
equipment
Immunity from
power frequency
magnetic field
Standard
Class/
Severity Level
CISPR 11
Group 1 Class B
CISPR 11
Group 1 Class B
IEC 61000-3-2
AC mains
IEC 61000-3-3
AC mains
IEC 61000-1-2
8 kV contact
discharges & 15kV
air discharges
IEC 61000-4-3
3.0; 10.0 V/m; 80
MHz + 2.7 GHz,
80% AM, 1kHz
IEC 61000-4-3
List of frequencies,
from 9 V/m up to
28 V/m, PM (18 Hz
or 217 Hz), FM 1
kHz
IEC 61000-4-8
30 A/m @ 50Hz &
60Hz
27
EN
Test
Result
N/A
Complies
N/A
N/A
Complies
Complies
Complies
Complies