Instructions for Use
Use
Use inert gas free from dust, particles or humidity (like Nitrogen) for venting the
CAUTION!
pump. The pressure at the vent port must be less than 2 bar (above atmospheric
pressure).
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal. Do not
use the pumping system in presence of explosive gases

Switching on and Use of TPS-mobile

328/400
After TPS-mobile unpacking, perform a brief visual analysis of
the system to be sure that no sign of damages due to
transportation is present (critical parts are: TMP flange,
Forepump fastening brackets, system front panel and the system
wheels).
Check the system voltage on the Agilent label stuck on the system
electronic units case.
TPS-mobile User Manual / 87-901-005-01 (F)