Switching On And Use Of Tps-Mini - Agilent Technologies TPS-mini Serie Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 61
16
Instructions for Use
Use
Use inert gas free from dust, particles or humidity (like Nitrogen) for venting the
CAUTION!
pump. The pressure at the vent port must be less than 2 bar (above atmospheric
pressure).
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal.
Do not use the pump in presence of explosive gases.

Switching on and Use of TPS-mini

To switch on the TPS-mini it is sufficient to supply the mains and
then move the external switch to ON position (red light on). The
integrated controller automatically recognizes the mains presence
and start up the pump.
At the first start up it is recommended to use the "Soft Start" mode by
enabling it on the controller. For the following start ups it is
recommended to disable the "Soft Start" mode. For the "Soft Start"
mode activation procedure, see the paragraph "Signal Description" in
the chapter "Technical Information".
262/318
TPS-mini User Manual / 87-901-003-01 (E)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969918096991819699182

Tabla de contenido