Vent Valve
CLOSED
OPEN
STOP
START COMMAND
Pump Connector
A 30-cm cable is available to connect the controller to the
pump. The following figure shows the controller output con-
nector configuration with pins:
•
A-F = pump temperature sensor
•
B-C-D = 60 Vac 3-phase output to pump motor stator
•
E = EARTH PROTECTION
Controller-to-Pump Connector
If the temperature sensor is disconnected, the pump will not
start. To prevent damage to the pump when the temperature
exceeds 60° C, the sensor automatically cuts off the power
supply
RAMP
NORMAL
STOP COMMAND
Vent Valve diagram in "Auto" mode
Vent Valve Delay
Vent Valve Open
Vent Valve Relay
Win 126
Time Win. 147
Win. 126
J5 – GAUGE CONNECTOR
This connector is used to connect the "EyeSys Mini-IMG"
gauge with its extension cable (see orderable parts table).
The pressure can be read via serial line (WIN. 224) or by RE-
MOTE I/O on Programmable Analog Out (in this case, the
controller replicates the signal coming from the gauge).
The Mini-IMG gauge is intended for high vacuum use; there-
fore, with a pressure above 50 mTorr, the signal is misinter-
preted and a "false" pressure is displayed. Use a transition
gauge to check whether the reading is effectively below 50
mTorr.
40/50
TECHNICAL INFORMATION
J5 Gauge Connector
NOTE
87-901-004-01 (B)