Agilent Technologies TwisTorr 84 FS AG Serie Manual De Instrucciones página 284

Ocultar thumbs Ver también para TwisTorr 84 FS AG Serie:
Tabla de contenido

Publicidad

17
Technical Information
Profibus Option
5
1
-
6
1
-
284/300
-
Error Code (bit field)
B
– Too High Load (the gas load is too high for the
7
pump)
B
– Short Circuit (a short circuit happened between
6
the two of the three motor phases)
B
– Over Voltage (too high bus voltage during
5
braking)
B
– Verify Line Select
4
Not used
B
– Power Fail (internal circuitry failure)
3
B
– Controller Over-Temperature ("Controller
2
Power Section Temperature" or "Controller CPU
Section Temperature" are over the limit)
B
- Pump Over-Temperature ("Pump Bearing
1
Temperature" or "Pump Body Temperature" are
over the limit)
B
– Check Connection To Pump (the connection
0
between controller and pump missing)
-
B
– Set Point R1 Status
0
0 = open
1 = closed
B
– Set Point R2 Status
1
0 = open
1 = closed
B
– Vent Valve Status
2
0 = closed
1 = open
B
– Not Used
4
B
– Not Used
5
B
– Not Used
6
B
– Not Used
7
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (D)
206
221
225
122

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X3508-64001X3508-64002

Tabla de contenido