Output Buffer - Agilent Technologies TwisTorr 84 FS AG Serie Manual De Instrucciones

Ocultar thumbs Ver también para TwisTorr 84 FS AG Serie:
Tabla de contenido

Publicidad

17
Technical Information
Profibus Option

Output Buffer

Tab. 20 Output buffer
Offset
Size
Unit Res
Byte
Byte
0
1
1
2
Hz
3
2
Hz
282/300
Range
Description
B
= Start/Stop
7
0 = Stop, 1 = Start
B
= Low Speed
6
This bit set the pump target speed to "low speed
value" (Output Buffer, Offset 3), otherwise the
target speed is set to "high speed value" (Output
Buffer, Offset 1).
0 =Off, 1 =On
B
– Soft Start
5
This bit activates the soft start function which run-
up the pump smoothly. Please activate this function
only if the pump has been stopped for at least3
months.
0 = On, 1= Off
B
= Vent Valve Command
4
0 = Close, 1 = Open
Valid only if in Par Buff, Offset-20, B
B
= Purge Valve Command
3
0 = Close, 1 = Open
Valid only if Par Buff, Offset-20, B
B
– Not used
2
B
= Operating Mode
0-1
These bits define from which source the controller
has to get commands (readings are always
available). In order to operate the controller by
Profibus these bits must be set to 3.
0 = Remote I/O
1 = Serial
2 = Front panel
3 = Profibus
1100 -
High Speed Value
1350
Set the high rotational speed value
1100 -
Low Speed Value
1350
Set the low rotational speed value
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (D)
WIN
Serial
000
001
100
122
=1
0
145
=1
0
008
120
117

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X3508-64001X3508-64002

Tabla de contenido