Technical Information
Controller Connection
J7 – Vent
Figure 7
This is dedicated 24 Vdc connector to control the optional vent valve
(see orderable parts table). The vent valve can be driven by controller
automatically or by serial line.
If you already have a vent valve installed 110 Vac powered, this valve
must be replaced by a new model (see orderable parts table).
Vent Valve
CLOSED
Figure 8
214/294
Vent Connector
OPEN
STOP
RAMP
START COMMAND
Vent Valve diagram in "Auto" mode
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01
Vent Valve Delay
Vent Valve Open
NORMAL
Win. 126
STOP COMMAND
Time Win. 147
(F)