Technical Information
Controller Connection
Pins
Name
5-com
Soft Start
The following figure shows a typical contact logic input connection
and the related simplified circuit of the controller.
Figure 10
J1 – Output
Figure 11
216/294
Description
Contact closed= Active
Contact open= Inactive
If active before the start command
activates the spft start procedure. To
be used only at first pump run-up.
Typical logic input connection
Logic output connector
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01
Electrical
Note
Optically isolated
Active only in
Vlow, max = 10v
Remote Mode
Imax = 5mA
operation, and
Imin (on) = 2mA
in STOP status.
(F)