When no external input-output device is available this connector
must be closed with the supplied mating connector that short-circuits
the START and INTERLOCK inputs with the GROUND input (see the
following figure).
Figure 18
How to Connect the Output Open Collector of the
Controller
Here below there are examples of typical connections of the input of
TV 401/301 MS to an external system. Two cases are considered:
1
2
Please note that on the connector a 24 Vdc, 60 mA voltage, a
GROUND signal and the open collector pin are available.
The pin numbers reported in the drawings are referred to the
"START/STOP" input, but the connections could be extended to all
the other controller inputs.
TV
401/301 MS
User Manual / 87-901-013-01
the customer supplies the 24 Vdc
the customer does not supply the 24 Vdc.
Technical Information
(A)
17
J5 - In-Out
223/264