TV 401/301 MS Controller Installation

The 969-8979M005 controller can not be mounted on bottom of the
pump, like the standard controller, because it is not supplied with
the plastic pads for snap mounting. It can be mounted on any flat
surface by means of the three threaded holes on the controller casing
(see controller outline drawing).

Connection of the Fore-Vacuum Pump

A KF 16 flange is available to connect the TV 401/301 MS pump to the
fore-vacuum pump. A vacuum approved pipe can be used. If a rigid
pipe is used, any vibration generated by the mechanical pump must
be eliminated through the use of bellows.
The TV 401/301 MS pump is characterized by its high compression ratio also for
NOTE
oil vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil backstreaming.
TV
401/301 MS
User Manual / 87-901-013-01
Technical Information

TV 401/301 MS Controller Installation

(A)
17
245/264