Descargar Imprimir esta página

Agilent Technologies TwisTorr Medium-TMP Manual De Instrucciones página 160

Ocultar thumbs Ver también para TwisTorr Medium-TMP:

Publicidad

17
Technical Information
Procedure to Connect the Serial and I/O Ports to an External
Cable
When a data is indicated between two quotes ('...') it means that the indicated
NOTE
data is the corresponding ASCII character.
Data type
Logic (L)
Numeric (N)
Alphanumeric (A)
The addressed SLAVE will respond with an ANSWER whose structure
depends from the MESSAGE type.
When the MESSAGE is a reading command, the SLAVE will respond
transmitting a string with the same structure of the MESSAGE.
160/181
<STX> (Start of transmission) = 0x02
<ADDR> (Unit address) = 0x80 (for RS 232)
<ADDR> (Unit address) = 0x80 + device number (0 to 31) (for RS
485)
<WIN> (Window) = a string of 3 numeric character indicating the
window number (from '000' to '999'); for the meaning of each
window see the relevant paragraph.
<COM> (Command) = 0x30 to read the window, 0x31 to write into
the window
<DATA> = an alphanumeric ASCII string with the data to be
written into the window. In case of a reading command this field
is not present. The field length is variable according to the data
type as per the following table:
Field length
1
6
Up to 14
<ETX> (End of transmission) = 0x03
<CRC> = XOR of all characters subsequent to <STX> and
including the <ETX> terminator. The value is hexadecimal coded
and indicated by two ASCII character.
TwisTorr Medium-TMP on-board Controller User Manual / 87-901-050-01
Valid characters
'0' = off
'1' = on
'-', '.', '0' . . . '9' right
justified with '0'
from blank to '_' (ASCII)