The kit is assembled as shown in the figure.
Figure 37
Two rubber or plastic hoses from the water supply must be fitted to
the two water cooling banjos.
Cooling may be carried out either through an open circuit with
eventual discharge of the water, or using a closed circuit cooling
system.
The water temperature must be between +10 °C and +30 °C, with an
inlet pressure between 3 and 5 bar. This allows a flow of about
200 l/h.
The water electrical conductance must be ≤ 500 μs/cm. When the
NOTE
conductance is higher, in closed water circuit, the use of up to 20 % of Ethyl-
Glycole is suggested.
TV 551/701 SEM Navigator User Manual / 87-900-958-01(G)
Model 969-9347
Technical Information
Water Cooling Kit Installation
17